products details

Created with Pixso. Home Created with Pixso. Products Created with Pixso.
Oil Testing Equipment
Created with Pixso.

Transformer Oil Testing Equipment SH121Automatic degassing shaker gas chromatography method

Transformer Oil Testing Equipment SH121Automatic degassing shaker gas chromatography method

Brand Name: Shengtai Instrument
Model Number: SH121
MOQ: 1
Payment Terms: T/T
Supply Ability: 10000sets/year
Detail Information
Place of Origin:
China
Display:
Large Screen LCD Display (240*128)
Temperature Range:
Room Temperature ~100℃
Power Source:
AC 220V±22V
Temperature Control Accuracy:
±0.2℃
Oscillation Frequency:
275±3 Times/min
Oscillation Amplitude:
35mm
Equation 1:
After Constant Temperature At 50℃ For 10 Minutes, Oscillate For 20 Minutes And Rest For 10 Minutes.
Formula 2:
After Constant Temperature At 50℃ For 10 Minutes, Oscillate For 5 Minutes And Keep The Constant Temperature At 75℃
Self-defined Meaning:
20~99 Minutes To Set The Oscillation, Static Time; Set Arbitrary Temperature Within 20~99℃
Sample Size:
Four 250ml Triangular Flasks Can Be Placed 8 100ml Syringes Can Be Placed
Instrument Size:
500mm × 350mm × 375mm
Net Weight:
31.5kg
Packaging Details:
wooden case ;Package size: 580mm × 430mm × 465mm gross weight: 39kg
Highlight:

automatic degassing shaker transformer oil testing

,

gas chromatography transformer oil testing equipment

,

SH121 transformer oil degassing shaker

Product Description

SH121 Degassing vibration instrument

The SH121 automatic degassing oscillator complies with the GB/T17623 DL429.4 standard and is used for oscillating degassing in the gas chromatography method for determining the content of dissolved gas components in oil; The oscillation degassing method for the determination of water-soluble acids in oil (colorimetric method) can also be used for constant temperature and timed oscillation in other physical and chemical experiments.

  • GB/T7598-2008 Determination of water-soluble acids in transformer oil during operation
  • GB/T17623-2017 Gas chromatography method for the determination of dissolved gas components in insulating oil;
  • GB/T259-88 Petroleum products - Determination of water-soluble acids and bases.

Main features

  • Adopt blue LCD large screen display, Chinese menu, no logo button.
  • Sliding door design makes it more convenient to take and place samples.
  • It has the characteristics of small volume, low noise, uniform distribution of temperature field and eight samples oscillating at a time.
  • The working process is fully automated, which reduces the labor intensity of the staff and improves the work efficiency. It is an ideal replacement product for the laboratory

Technical parameters

Display large screen LCD display (240*128);
Temperature range room temperature ~100℃;
Power source AC 220V±22V
Temperature control accuracy

room temperature ~50℃±0.2℃; 50℃~100℃±0.3℃

 

Oscillation frequency 275±3 times/min;
Oscillation amplitude 35mm
Equation 1 After constant temperature at 50℃ for 10 minutes, oscillate for 20 minutes and rest for 10 minutes.
Formula 2 after constant temperature at 50℃ for 10 minutes, oscillate for 5 minutes and keep the constant temperature at 75℃;
Self-defined meaning

20~99 minutes to set the oscillation, static time;

Set arbitrary temperature within 20~99℃

Sample size

Four 250ml triangular flasks can be placed

Eight 100ml syringes can be placed;

Instrument size: 500mm × 350mm × 375mm net weight: 31.5kg
Package size: 580mm × 430mm × 465mm gross weight: 39kg

 

Transformer Oil Testing Equipment SH121Automatic degassing shaker gas chromatography method

Packing list

S/N Name Quantity Unit Remarks
1 Host 1 set  
2 Oscillating disk 1 set  
3 power cord 1 piece  
4 Instructions 1 copy  
5 Packing list 1 copy  