Détails des produits

Created with Pixso. Accueil Created with Pixso. Produits Created with Pixso.
Appareil de contrôle de point d'inflammabilité
Created with Pixso.

Transformer Oil Cold Pressed Oil Flash Point SD261 Closed Cup Flash Point Apparatus

Transformer Oil Cold Pressed Oil Flash Point SD261 Closed Cup Flash Point Apparatus

Nom De La Marque: Shengtai Instrument
Numéro De Modèle: SD261
MOQ: 1
Conditions De Paiement: T/T
Capacité D'offre: 10000sets / an
Des informations détaillées
Lieu d'origine:
Chine
Alimentation de travail:
AC220V±10%, 50Hz.
Dispositif de chauffage:
puissance de chauffage de 0 à 600 W réglable en continu
Le taux de chauffage témoin:
1 | 12℃/mn.
Tasse d'huile standard:
Diamètre intérieur: 50,7 ~ 50,8 mm. Profondeur: 55,7 ~ 56,0 mm. Capacité de test Profondeur d'en
Dispositif d'allumage:
Source d'allumage: ouverture d'arme à feu à gaz: 0,8 mm
Thermomètre:
Thermomètre à barres de mercure, conforme aux exigences de la norme GB/T 261-2008.
Détails d'emballage:
caisse en bois
Mettre en évidence:

Transformer oil flash point tester

,

Cold pressed oil flash tester

,

SD261 closed cup flash apparatus

Description du produit

Diesel oil above 40 ℃, transformer oil, cold pressed oil flash point SD261 closed flash point meter

Transformer Oil Cold Pressed Oil Flash Point SD261 Closed Cup Flash Point Apparatus 0

SD261 closed mouth flash point meter is designed and manufactured in accordance with the requirements of the national standard GB/T 261-2008 "determination of flash point Binsky-Martin closed mouth cup method". It is suitable for the determination of the flash point of petroleum product samples with a flash point above 40℃ by the method specified in the GB/T 261-2008 standard.

 

GB/T 261-2008manual Binsky-Martin closed cup flash point tester for laboratory testing

 

GB/T 261-2008manual Binsky-Martin closed cup flash point tester for oil testing

 

Portable Manual Binsky-Martin Closed Cup Flash Point Tester Price

 

Product model SH105B
Test sample Petroleum, electric power, chemicals, products
Compliance with standards GB261-2008 ASTM, D93 Conforming to ISO-2719, ASTM D93, Pensky-Marten closed cup method
Simple technical features or parameters Air source ignition, manual timing and visual observation

 

I. Main technical features

Heating power continuous, stepless adjustable.

II. Main technical parameters and indicators

working power supply AC220V±10%, 50Hz.
Heating device heating power from 0 ~ 600W continuously adjustable.
The sample heating rate 1 ~ 12℃/ min.
Standard oil cup

Inner diameter: 50.7 ~ 50.8 mm.

Depth: 55.7 ~ 56.0 mm.

Test capacity notch depth: 33.9 ~ 34.3 mm.

Test capacity: about 70 ml

Ignition device

ignition source: gas

Firearm aperture: 0.8mm

Thermometer bar mercury thermometer, in line with GB/T 261-2008 standard requirements.

Packing list:

The host 1
Oil cup 1
Thermometer 1
Instructions 1
Certificate warranty card 1

 