How to Determine Water-Soluble Acid of Fire-Resistant Hydraulic Fluid
Fire-resistant hydraulic fluid is phosphate ester synthetic fire-resistant hydraulic fluid (not mineral oil). It is a special hydraulic medium for the EH electro-hydraulic control system of steam turbines in power plants. It transmits pressure and drives servo valves. Its key feature is an extremely high autoignition point; it is not easy to ignite and can self-extinguish once on fire. The mainstream type is triaryl phosphate ester, a pale yellow transparent liquid with density>1 (approximately 1.13~1.17 g/cm³). Its autoignition point is generally above 530℃, much higher than ordinary turbine oil. Leakage at high temperature will not easily ignite when contacting hot pipelines, ensuring unit safety.
Test Purpose
Judge hydrolysis and early deterioration of oil: Detection of water-soluble acid (pH drop) indicates that the fire-resistant hydraulic fluid has undergone hydrolysis or oxidation. It serves as an early warning indicator of oil degradation and can identify deterioration tendency earlier than total acid value.
Evaluate corrosion risk of the system: Accumulated water-soluble acid together with moisture in the system will corrode servo valves, metal pipelines and seals of the EH system, resulting in valve core sticking and internal leakage, which directly affects the stable operation of the steam turbine control system.
Acceptance of new oil and regular monitoring of in-service oil: Water-soluble acid is not permitted in new oil. The pH value of in-service oil shall be monitored periodically to guide bypass regeneration, dehydration filtration or oil replacement. This is one of the routine monitoring items for fire-resistant hydraulic fluid in power plants.
Assist in identifying pollution sources: Distinguish whether the oil deterioration is caused by self-hydrolysis and aging or external acidic contaminants.
Test Sample: Fire-resistant hydraulic fluid
Test Instrument: SH259B Automatic Water-Soluble Acid Tester, compliant with GB/T 259
![]()
Test Procedure
Sampling: Measure 50 mL of sample and 50 mL of neutral distilled water, and transfer both into a 250 mL separatory funnel. Fire-resistant hydraulic fluid is viscous, so heat the mixture in a water bath to 50~60 ℃ (light oils such as gasoline are not heated). Cap the funnel and release pressure appropriately during operation.
Extraction: Shake steadily for 5 minutes to fully contact and extract the oil phase and aqueous phase, then stand for stratification to avoid emulsification. If emulsification occurs, replace distilled water with 1:1 ethanol-water solution at 50~60 ℃.
Separation and filtration: After clear stratification, slowly drain the lower aqueous extract and filter it through qualitative filter paper into a clean conical flask.
Measurement (two methods; the pH meter method is preferred for arbitration):
Indicator method: Take two portions of filtrate. Add methyl orange indicator to one portion and phenolphthalein indicator to the other, then observe color change.
pH meter method: Measure the pH value of the extract with a calibrated pH meter. Run the test in duplicate and take the average value as the result. The difference between the two results shall be ≤0.05 pH to be acceptable.
Blank test: Carry out a blank test with distilled water simultaneously.
Test Result
Judgment by Indicator
Methyl orange turns red: Water-soluble acid detected
Phenolphthalein turns red: Water-soluble alkali detected
No color change for both indicators: No water-soluble acid or alkali detected
Judgment by pH Meter Method (Recommended)
Record the pH value of the extract. New fire-resistant hydraulic fluid is generally required to be free of water-soluble acid (pH close to neutral 7). A low pH value means the oil has hydrolyzed and deteriorated, generating water-soluble acidic substances that will corrode EH servo valves, pipelines and seals. Oil filtration, regeneration or replacement is recommended.
Precision Judgment
The test result is valid if the pH difference between two parallel determinations is ≤0.05. Retest if the difference exceeds the limit.